UNSWNuclear Magnetic Resonance Facility

CONTACTS | HOME

ANALYTICAL CENTRE

NMR

Nuclear Magnetic Resonance Facility

Quick estimation of T1s (Derome p. 166)

Many 2D experiments require a knowledge of the T1s of the nuclei involved for optimum conditions to be set up.

A ball park figure (usually all that's needed) can be estimated from an inversion recovery sequence:

Inversion Sequence

The Δ delay is manually adjusted until the signal(s) of interest are nulled in the resulting spectrum. (If Δ short relative to T1, signals negative; if Δ long relative to T1 signals positive).

It can be shown that Δ delay for null = T1ln2 i.e. T1 = 1.44*Δnull.

For protons, usually only one scan is required for each attempt to find the null.

Care must be taken to ensure that at least 5T1s are left between successive attempts to find the null point (or scans if more than one needed) and that the same phase correction is always applied to determine sign of the signals.

Prior calibration of the 90° pulse is also a good idea so that an accurate 180° pulse can be applied (probably will need to be known for the 2D experiment also).

Numerous reasons exist for wanting to know the approximate T1.

e.g. Optimum repetition rate for best sensitivity is 1.3 T1 usually - useful for simple COSY experiments.

NOESY on small molecules requires the mixing time to be set to the order of the T1s of the protons of interest. Where there is a range of T1s, a mixing time of the shortest T1 is often most useful. A recycle delay time of 3 times the longest T1 is often a good compromise.

Inverse detected proton carbon correlations (i.e. proton observation) require the repetition time to be 1.3-2 times the shortest proton T1, for fast acquisition.

The appropriate pulse programs is quickt1 on the DMX/DPX. This is most easily done using the quickt1 macro in XWINMMR:

Procedure:

  1. Run a normal 1H spectrum then calibrate the 90° pulse
  2. Type "quickt1". This loads the pulse program then requests some parameters:
    pl1 - power level for pulse. This will be the same if the same dataset as the pulse calibration is used
    p1 - value for 90° pulse calibrated in 1).
    d7 - d7 corresponds to the Δ delay. Start with a small value, e.g., 100μs
  3. Run a 1 scan spectrum and process with fp or efp. For a short D7 all peaks should be negative but in phase.
  4. Repeat with various values of D7 so that peaks of interest are nulled (zero). Ignore solvent and impurity peaks. D7 = 1s is a good first guess.
  5. Remember T1 = 1.44*D7. This rough method is most likely to understimate the T1 by a few percent.
Back to Local Information

NMR

Mail Address:


NMR Facility
UNSW Analytical Centre
Basement Level, Room B41
Chemical Sciences Bldg F10
University of NSW
Kensington NSW 2033


T +61 (2) 9385 4705
F +61 (2) 9385 4699

 

Deliveries (before 4pm):


Chemistry Store, S108
Chemical Sciences Bldg F10
Via Gate 2 High Street
University of NSW
Kensington NSW 2033

nmr@unsw.edu.au

Location Maps & Deliveries

Privacy Policy     Copyright & Disclaimer     Accessibility     email Enquiry

NMR Facility- UNSW - Chemical Sciences Building F10, Sydney, NSW 2052 Australia | Tel: +61 (2) 9385 4705 Fax: +61 (2) 9385 4699
© Copyright 2007 UNSW ANALYTICAL CENTRE | CRICOS Provider Code: 00098G | Authorized by the NMR Facility Manager
Page Last Updated: Wednesday, June 25, 2008 1:13 PM